Cat.No. |
R11-650-025 |
Model |
RTS-8 plus |
Light source |
Laser |
Measurement wavelength (¥ë) |
850 ¡¾ 15 nm |
Measurement range |
0-100 OD600 |
E.coli factory calibration measurement range |
0-50 OD600 |
S.cerevisiae factory calibration measurement range |
0-75 OD600 |
Achievable user calibration measurement error (range
0.1-6 OD600) |
¡¾ 0.3 |
Achievable user calibration measurement error (range
6-50 OD600) |
¡Â 5% |
Achievable user calibration measurement error (range
50-75 OD600) |
¡Â 10% |
Measurement periodicity per hour |
1-60 |
Temperature setting range |
+15¡É ... +60¡É |
Temperature control range |
+15 ¡É below ambient ... +60 ¡É |
Temperature stability |
¡¾0.3 ¡É |
Sample temperature accuracy (20¡É
-37¡É) |
¡¾>1¡É |
Tube sockets |
8 |
Sample working volume range |
3-50 ml |
Speed control range |
150-2700 rpm |
Reverse spin time setting range 150-250 rpm |
0 s |
Reverse spin time setting range 250-300
rpm |
2-60 s |
Reverse spin time setting range 300-2700
rpm
| 0-60 s |
Display |
LCD |
Minimum PC requirements |
Intel/AMD Processor, 1 GB RAM, Windows
Vista/7/8/8.1/10, USB 2.0 port |
Dimensions (W¡¿D¡¿H) |
350¡¿690¡¿300 mm |
Weight |
20 kg |
Nominal operating voltage |
AC 230 V, 50 Hz |
Power consumption |
3.15 A / 500 W |
** |
** |
02 sensor |
+ |
Range |
0-100% |
Accuracy |
¡¾0.05% O2 at 0.2%,
¡¾0.4%
O2 at 20.9% |
Drift |
<0.03% O2 within 30 days |
Temperature range |
up to 40¡É |
Response time (t90) |
<6 s |
Storage stability |
18 months |
** |
** |
pH sensor |
+ |
Range |
4.0 - 8.5 pH |
Accuracy |
¡¾0.10 pH at pH 7 |
Drift |
<0.005 pH per day |
Temperature range |
up to 40¡É |
Response time (t90) |
<120 s |
Storage stability |
18 months |