Cat. No. | Model | Spectral Bandwidth |
---|---|---|
S23-3269-041 | OPTIZEN Alpha | 1 nm (190 to 1100 nm) |
Cat. No. | S23-3269-041 |
---|---|
Model | OPTIZEN Alpha |
Photometrics System | Double-beam type |
Light Source | Tungsten Halogen Lamp & Deuterium Lamp (Built-in light source auto interchanging motor) |
Spectral Bandwidth | 1 nm (190 to 1100 nm) |
Wavelength Range | 190 to 1100 nm |
Wavelength Display(setting) | 0.05 nm |
Wavelength Accuracy | ¡¾ 0.3 nm (For entire range)
¡¾ 0.1 nm (at 656.1 nm) |
Wavelength Repeatability | <¡¾ 0.1 nm |
Slew Rate | About 13,500 nm/min |
Scan Speed | max 6,000 nm/min |
Photometric Range | Absorbance : -4 to 4 Abs Transmittance : 0 % to 400 % |
Photometric Accuracy | ¡¾0.002 at 0.5 Abs ¡¾0.004 at 1.0 Abs ¡¾0.006 at 2.0Abs |
Photometric Repeatability | ¡¾ 0.0002 at 0.5 Abs ¡¾ 0.0006 at 1.0 Abs ¡¾ 0.001 at 2.0 Abs |
Stray Light | < 0.02 % Nal at 220 nm, NaNO©ü at 340 nm |
Baseline Stability | < 0.0003 Abs/h |
Baseline Flatness | <¡¾ 0.0005 Abs |
Monochromator | Czerny-Turner type with 1200 lines/nm blazed grating |
Lamp Interchange Wavelength | 340~410 nm (Default 370 nm) |
Standard Cell Holder | Automatic Rotary type 8-position Multi-Cell Holder |
Display | 8 inch color LCD with touch screen |
Operating System(OS) | Windows 10 (Embedded PC) |
Power Requirement | 100~240 V, 50~60 Hz |
Weight | 14 kg |
Dimensions(W£øD£øH) | 520 mm£ø500 mm£ø200 mm |